Biochips based on semiconductor and manufacturing expertise
ROHM biochips utilize a combination of microfluidics, microfabrication, design, and optical technologies. Microchannels allow chemical reactions to occur in an extremely small space via reagents.
Next, sensing technology enables accurate detection of target biomarkers using a trace amount of blood. Biochips such as these are expected to play a major role in health care and medical testing in the future.
B-Analyst realizes rapid and high accuracy measurement with only a drop of blood by utilizing μTAS and liquid reagents.
- Simple Operation
B-analyst performs measurement from plasma separation to data output with a drop of blood in 7min 30sec. The measurement is completed by following steps
- High Accuracy
High correlation to reference methods.
Reliable measurement result can be provided.
- Easy Maintenance
Cleaning and washing are not required.
Calibration curve data corrected via 2D barcode.
SPECIFICATIONS
Instrument specifications
Detection wavelength | 635 nm | |
Detection stability | ±0.001 Abs. | |
Incubation temperature | 37.0 ± 1.0℃ | |
Interface | For data input – output x 1 For bar code reader connection x 1 |
|
Display | 20 rows x 4 columns monochrome LCD | |
Printer | Built-in thermal printer (paper width: 58 mm) | |
Data memory | For recording measurement results of latest 100 samples | |
Calendar clock | Built-in (backup battery CR2032) | |
Electrical characteristic | Voltage | AC100~240 V |
Supply frequency | 50/60 Hz | |
Power input | 250 W | |
Condition of use | Temperature | 15~30℃ |
Humidity | 30~80% | |
Dimension / Weight | 340(W)×290(D)×270(H) mm / 9.5 kg |
Reagent Chip specifications
Product name | B-analyst HbA1c | B-analyst CRP | B-analyst hsCRP |
Sample | Whole blood | Whole blood · Plasma · Serum |
Whole blood · Plasma · Serum |
Sample volume | 4 μL | 4 μL | 9.5 μL |
Measurement method | Latex agglutination immunology turbidimetric method | ||
Measurement time | 7’30” | ||
Measurement range | 3.3 – 12.6 %![]() |
0.1 – 20.0 mg/dL![]() |
0.01 – 4.00 mg/dL![]() |
Size
For inquiries about this product:
Contact: Mark Augustus M. Tecson
Cell: 0917-866-5767
Landline: 02 515-5548
Email: mark.tecson@twinj3.com